Remote Control Commands
R&S
®
NRX
192
User Manual 1178.5566.02 ─ 05
<undef>
1 to n
No suffix required.
Usage:
Event
CALCulate<Measurement>:LIMit<undef>:FAIL?
Queries whether upper or lower limits have been exceeded. The status is reset if one
of the following events occurs:
●
Device is switched on.
●
).
●
CALCulate<Measurement>:LIMit<undef>:CLEar[:IMMediate]
ted.
Suffix:
<Measurement>
.
1 to 4
Measurement channel
<undef>
1 to n
No suffix required.
Usage:
Query only
CALCulate<Measurement>:LIMit<undef>:FCOunt?
Queries the number of limit violations that occurred. The counter is zeroed if one of the
following events occurs:
●
Device is switched on.
●
).
●
CALCulate<Measurement>:LIMit<undef>:CLEar[:IMMediate]
ted.
Suffix:
<Measurement>
.
1 to 4
Measurement channel
<undef>
1 to n
No suffix required.
Usage:
Query only
CALCulate<Measurement>:LIMit<DirectionalChannel>:LOWer:STATe
<state>
Enables or disables the checking of the lower limit.
Suffix:
<Measurement>
.
1 to 4
Measurement channel
Measurement Settings and Results