Sensor Configuration
R&S
®
NRX
122
User Manual 1178.5566.02 ─ 05
"Full"
The maximum bandwidth of the sensor is used.
Remote command:
CALCulate<Measurement>[:CHANnel<Channel>]:NRT:VBWidth[:VALue]
[SENSe<Sensor>:]BANDwidth:VIDeo:FNUMber
[SENSe<Sensor>:]BWIDth:VIDeo:FNUMber
Integration Time Mode
Specifies which integration time is used for a single measurement.
"Default"
Uses the default settings.
"User"
Define a value under
.
Remote command:
CALCulate<Measurement>[:CHANnel<Channel>]:NRT:APERture:MODE
Integration Time
Available if
is set to "User".
Defines the integration time for a single measurement.
Remote command:
CALCulate<Measurement>[:CHANnel<Channel>]:NRT:APERture[:VALue]
Averaging Mode
Sets the averaging mode.
"User"
Define the value under
.
"Auto"
Determines the average count automatically from the level of the
input signal.
Remote command:
CALCulate<Measurement>[:CHANnel<Channel>]:AVERage:COUNt:AUTO[:
STATe]
Averaging Count
Available if
is set to "User".
Sets the number of readings that are averaged for one measured value. The higher the
count, the lower the noise, and the longer it takes to obtain a measured value.
Remote command:
CALCulate<Measurement>[:CHANnel<Channel>]:AVERage:COUNt[:VALue]
For NRT Measurement Type