Sensor Configuration
R&S
®
NRX
114
User Manual 1178.5566.02 ─ 05
Remote command:
CALCulate<Measurement>[:CHANnel<Channel>]:AVERage:COUNt:AUTO:
MTIMe
Timeslot
Available for continuous average, burst average, time gate measurements.
Sets a timeslot from which the measured value is used to determine the filter length
automatically. The timeslot number must not exceed the number of the currently set
timeslots.
Remote command:
CALCulate<Measurement>[:CHANnel<Channel>]:AVERage:COUNt:AUTO:
SLOT
Moving Average State
Available for continuous average, burst average, trace, pulse analysis, timeslot, statis-
tics measurements.
Enables or disables the automatic termination control.
Remote command:
CALCulate<Measurement>[:CHANnel<Channel>]:AVERage:TCONtrol:AUTO
CALCulate<Measurement>[:CHANnel<Channel>]:TRACe:AVERage:
TCONtrol:AUTO
Moving Average
Available for continuous average, burst average, trace, pulse analysis, timeslot, statis-
tics measurements.
Defines how the measurement results are output. This is called termination control.
On
Outputs intermediate values to facilitate early detection of changes in
the measured quantity. In the settled state, that means when the
number of measurements specified by the average count has been
performed, a moving average is output.
Off
Specifies that a measurement result is not output until the entire mea-
surement has been completed. This means that the number of mea-
surement cycle repetitions is equal to the set average count. If the
average count is large, the measurement time can be very long.
Remote command:
CALCulate<Measurement>[:CHANnel<Channel>]:AVERage:TCONtrol[:
ENUM]
CALCulate<Measurement>[:CHANnel<Channel>]:TRACe:AVERage:
TCONtrol[:ENUM]
Averaging Domain
Requires an R&S
frequency selective power sensor.
Sets the averaging method. See also the user manual of the power sensor.
"Power"
Power averaging
"Video"
Logarithmic averaging