Remote Control Commands
R&S
®
NRX
318
User Manual 1178.5566.02 ─ 05
CALCulate<Measurement>[:CHANnel<Channel>]:TRACe:AVERage:COUNt[:
VALue]
<value>
Effective for trace measurements.
Sets the number of readings that are averaged for one measured value. The higher the
count, the lower the noise, and the longer it takes to obtain a measured value.
Suffix:
<Measurement>
.
1 to 4
Measurement channel
<Channel>
1 to 2
1 = primary sensor, 2 = secondary sensor
Parameters:
<value>
If the entered filter length is not a 2
n
value, the value is rounded
to the next 2
n
value without an error message.
Range:
1 to 65536
*RST:
4
Manual operation:
See
CALCulate<Measurement>[:CHANnel<Channel>]:TRACe:AVERage:TCONtrol:
AUTO
<state>
Effective for trace measurements.
Enables or disables the automatic termination control. See also
CALCulate<Measurement>[:CHANnel<Channel>]:TRACe:AVERage:
TCONtrol[:ENUM]
Suffix:
<Measurement>
.
1 to 4
Measurement channel
<Channel>
1 to 2
1 = primary sensor, 2 = secondary sensor
Parameters:
<state>
*RST:
ON
Manual operation:
See
CALCulate<Measurement>[:CHANnel<Channel>]:TRACe:AVERage:TCONtrol[:
ENUM]
<mode>
Effective for trace measurements.
Defines how the measurement results are output. This is called termination control.
Suffix:
<Measurement>
.
1 to 4
Measurement channel
Configuring Sensors