Remote Control Commands
R&S
®
NRX
265
User Manual 1178.5566.02 ─ 05
Query parameters:
<function>
For the description of the string, see
CALCulate<Measurement>[:CHANnel<Channel>]:
FEED<Channel>
Usage:
Query only
[SENSe<Sensor>:]POWer:REFLection:RANGe:AUTO
<state>
Enables or disables the automatic adaptation of the bargraph scaling for the reflection
indication. If enabled, the scale limits of the bargraphs are automatically adapted to the
current measured value.
Suffix:
<Sensor>
.
1 to 128
Configured sensor connected at: port A = 1, ... , port D = 4, R&S
NRX-B9 = 101, USB and LAN port = 5 to 100
Parameters:
<state>
ON | OFF
ON
The scale limits of the bargraphs are automatically adapted to
the current measured value.
OFF
The scale limits remain fixed.
*RST:
1
[SENSe<Sensor>:]POWer:REFLection:RANGe:LIMit:DETect
<value>
Effective if the Out
1
/
Trig
Out connector is configured as monitoring output for the
reflection indication, for example using
[SENSe<Sensor>:]POWer:REFLection:
ON
.
Defines when a logic high level (> 2.7
V) is output at the Out
1
/
Trig
Out connector.
Suffix:
<Sensor>
.
1 to 128
Configured sensor connected at: port A = 1, ... , port D = 4, R&S
NRX-B9 = 101, USB and LAN port = 5 to 100
Parameters:
<value>
INBound | OUTBound | HIGH
INBound
Measured power is within the range specified.
OUTBound
Measured power is out of the range defined.
HIGH
Measured power exceeds the upper scale limit.
*RST:
HIGH