Remote Control Commands
R&S
®
NRX
238
User Manual 1178.5566.02 ─ 05
CALCulate<Measurement>:TRACe:MEASurement:TRANsition:NEGative:
DURation?
Effective for pulse analysis measurements.
Queries the time the pulse requires to transition from the pulse top level to the pulse
base level.
Suffix:
<Measurement>
.
1 to 4
Measurement channel
Usage:
Query only
CALCulate<Measurement>:TRACe:MEASurement:TRANsition:NEGative:
OCCurrence?
Effective for pulse analysis measurements.
Queries the time when the signal passes through the medial reference power level with
falling edge, referenced to the delayed trigger event. Indicates the stop point of the first
power pulse within the analysis window.
Suffix:
<Measurement>
.
1 to 4
Measurement channel
Usage:
Query only
CALCulate<Measurement>:TRACe:MEASurement:TRANsition:NEGative:
OVERshoot?
Effective for pulse analysis measurements.
Queries the height of the local minimum before a rising edge, divided by the pulse
amplitude:
Negative overshoot = 100 % x
Pulse base power - minimum power
Pulse amplitude
DISPlay[:WINDow<Window>]:TRACe:
.
Suffix:
<Measurement>
.
1 to 4
Measurement channel
Usage:
Query only
CALCulate<Measurement>:TRACe:MEASurement:TRANsition:POSitive:
DURation?
Effective for pulse analysis measurements.
Measurement Settings and Results