Remote Control Commands
R&S
®
NRX
271
User Manual 1178.5566.02 ─ 05
Table 14-1: <expression> parameter
<expression>
Description
"(SENS1)"
Measured value of sensor A
"(SENS2)"
Measured value of sensor B
"(SENS3)"
Measured value of sensor C
"(SENS4)"
Measured value of sensor D
"(SENSn-SENSm)"
Difference between the measured values of sensor n and sensor m
"(SENSn+SENSm)"
Sum of the values measured by sensor n and sensor m
"(SENSn /SENSm)"
Quotient of the values measured by sensor n and m
"SWR(SENSn,SENSm)"
Standing wave ratio. The output unit is set to percent (
UNIT:RAT PCT
).
SENSn
SENSm
SENSn
SENSm
/
1
/
1
Sensor n measures the forward power of a wave, sensor m measures the
reflected power.
"REFL(SENSn,SENSm)"
Reflection coefficient/transmission factor of a DUT. The output unit is set to
percent (
UNIT:RAT PCT
).
SENSn
SENSm
/
Sensor n measures the forward power of a wave, sensor m measures the
reflected/transmitted power.
"RLOS(SENSn,SENSm)"
Return loss/transmission loss of a DUT. The output unit is set to dB
(
UNIT:RAT DB
).
This function principally supplies the same result as
"SENSn/SENSm"
. The
difference is that the output unit is automatically set to dB.
SENSn
SENSm
/
log
20
10
Sensor n measures the forward power of a wave, sensor m measures the
reflected/transmitted power.
With n and m = 1, 2, 3 or 4
Suffix:
<Measurement>
.
1 to 4
Measurement channel
Parameters:
<expression>
See
*RST:
Depends on the selected channel.
Default unit: Depends on <expression> and the set unit.
Manual operation:
See
"Channel Calculation Function"
Calculation Functions