Welcome to the R&S
ZNB/ZNBT
R&S
®
ZNB/ZNBT
20
User Manual 1173.9163.02 ─ 62
New "Reset" buttons in "Advanced Settings" dialogs.
●
External switch matrices: Support for R&S
OSP added
Solved issues
●
Remote automatic calibration: Visualization not shown
●
Calibration units R&S
ZV-Zxx not detected by 2nd generation R&S
ZNB
●
Help/manual: Wrong syntax description of remote command
[SENSe<Ch>:]SEGMent<Seg>:DEFine
●
Power calibration did not work correctly for segmented sweeps whose stimulus val-
ues are not monotonically increasing
●
R&S
ZNXSIM and R&S
ZNrun on the same PC: The firmware simulation did not
start if the R&S
ZNPC driver installed with R&S
ZNrun was newer than the driver
installed with the R&S
ZNA firmware. Error message: "No valid Partnumber found!"
●
Calibration kit Keysight 85058EP: wrong capacitance value C
0
in Open (f) circuit
model
●
Precision frequency reference R&S
ZNB/ZNBT-B4: No "OCXO Cold" warning if the
reference frequency quartz was not on working temperature
(Was only displayed with developer service level 3)
●
Time domain analysis R&S
ZNB/ZNBT-K2:
–
Selecting a different time domain transform ("Type" selection on "Time Domain"
tab) did not take effect
–
Changing the permittivity did not change the distance in TDR distance plots
●
R&S
ZNXSIM: The enhanced offline data analysis introduced in FW V3.40 was not
available in the R&S
ZNBT simulation
●
Distance to fault only worked for live traces
●
If Channel – [Channel
Config] > "Channels" > "Fixture Simulator" is enabled on a 2-
port R&S
ZNB, then the corresponding setup cannot be loaded into a 4-port
R&S
ZNB
Firmware version
► To check your R&S
ZNB/ZNBT firmware version, select "Help" > "About..." from the
main menu.
2.2
Documentation overview
This section provides an overview of the R&S
ZNB/ZNBT user documentation. Unless
specified otherwise, you find the documents on the R&S
ZNB/ZNBT product page at:
●
https://www.rohde-schwarz.com/manual/ZNB
●
https://www.rohde-schwarz.com/manual/ZNBT
Documentation overview