GUI reference
R&S
®
ZNB/ZNBT
540
User Manual 1173.9163.02 ─ 62
The exported ZIP file also contains the characterizations of the standards that were
used for calibration and verification. The characterization data are kept as static
resources in the METAS VNA Tools database. In summary, the exported ZIP file
includes everything that is required to perform an offline analysis using any METAS
VNA Tools installation, not only the one on the VNA.
Note:
The exported file can be directly unzipped to the METAS VNA Tools user data
folder (default:
C:\Users\Public\Documents\Metas.Vna.Tools
). After
(re-)opening the METAS VNA Tools, the exported project and its associated static data
are available for offline analysis.
METAS Reconnection
.
This button is only enabled, if a METAS cal is active.
5.11.5.2
METAS Calibration (Manual) dialog
This dialog allows you to set up and perform a
nel.
Access:
Channel – [Cal] > "METAS Cal" > METAS Calibration – "Start (Manual)"
Only a single VNA channel can be METAS calibrated at a time. If you have a METAS
calibration for another channel (no matter in which recall set), this METAS calibration is
deleted when you apply the new one.
Step 1: Ports and Type
In step 1, choose the ports to be calibrated and the calibration type to be performed.
Only calibration types supported by METAS VNA Tools are available.
Ports
Selects the test ports to be calibrated.
Cal softtool