GUI reference
R&S
®
ZNB/ZNBT
688
User Manual 1173.9163.02 ─ 62
Remote command:
[SENSe<Ch>:]HARMonic:DLENgth:DATA
[SENSe<Ch>:]HARMonic:ELENgth:DATA
[SENSe<Ch>:]HARMonic:MLENgth:DATA
Permittivity / Velocity Factor
Specifies the (relative) permittivity ε
r
of the DUT or, alternatively, its velocity factor 1/
sqrt(ε
r
).
The velocity factor is a measure for the velocity of an electromagnetic wave in a dielec-
tric with permittivity ε
r
, relative to its velocity in a vacuum (velocity factor < 1).
Permittivity and velocity factor are coupled parameters, i.e. setting one of them deter-
mines the other. A higher permittivity implies a smaller mechanical length, but leaves
delay and electrical length unchanged (see
Max. DUT Measure Delay / Max. El.
).
Remote command:
[SENSe<Ch>:]HARMonic:PERMittivity:DATA
[SENSe<Ch>:]HARMonic:VELocity:DATA
Rise Time Definition
Selects between rise time definitions 10%-90% (default) and 20%-80%.
Remote command:
[SENSe<Ch>:]HARMonic:RTIMe:THReshold
Rise Time
The minimum rise time (according to the selected
wishes to measure on the DUT.
The minimum rise time, in turn, determines the stop frequency of the frequency sweep:
the smaller the "Rise Time", the larger the stop frequency – limited by the analyzer's
maximum frequency.
Remote command:
[SENSe<Ch>:]HARMonic:RTIMe:DATA
TDR Stimulus - Advanced Settings dialog
The "TDR Stimulus - Advanced Settings " compiles advanced settings of the time
domain transformation that is available with the standard time domain option
R&S
ZNB/ZNBT-K2.
Access:
Applic softtool