Command reference
R&S
®
ZNB/ZNBT
1167
User Manual 1173.9163.02 ─ 62
MMEMory:STORe:CKIT:WLABel
<CalKitName>, <KitLabel>, <CalKitFile>
Stores the data of a calibration kit to a specified file. The calibration kit is identified by
its name and label.
Setting parameters:
<CalKitName>
Name of a user-defined calibration kit available on the analyzer.
Tip:
It is not possible to modify or store predefined or ideal kits.
<KitLabel>
Label of the calibration kit, usually its serial number.
<CalKitFile>
String parameter to specify the name and directory of the cal kit
file to be created. The file is a NWA-specific cal kit file with the
extension *.calkit.
If no path is specified the analyzer searches the current direc-
tory, to be queried with
Example:
See
[SENSe:]CORRection:CKIT:LLABel
Usage:
Setting only
MMEMory:STORe:CMAP
<ColorSchemeFile>
MMEMory:STORe:CMAP:HCOPy
<ColorSchemeFile>
Stores the user-defined color scheme (display/print) to a VNA color scheme file.
The
HCOPy
variant is for the print color scheme.
Setting parameters:
<ColorSchemeFile> String parameter to specify the name and directory of the color
scheme file to be created. If no path is specified the analyzer
uses the current directory, to be queried with
. The default extension (manual control) for color
scheme files is
*.ColorScheme
, although other extensions are
allowed.
Example:
See
Usage:
Setting only
Manual operation:
See
MMEMory:STORe:CORRection
<Channel>, <CalGroupFile>
Copies the correction data of channel
<Channel>
to the cal pool, generating a new
correction data file (cal group). The file has the extension
*.cal
and is stored in the
C:\Users\Public\Documents\Rohde-Schwarz\Vna\Calibration\Data
directory.
Setting parameters:
<Channel>
Channel number
SCPI command reference