Command reference
R&S
®
ZNB/ZNBT
865
User Manual 1173.9163.02 ─ 62
CALCulate<Chn>:EYE:INPut:OLEVel
<OneLevel>
CALCulate<Chn>:EYE:INPut:ZLEVel
<VoltageLevel>
Defines the highest/lowest (nominal) voltage level of the simulated digital signal gener-
ating the related eye diagram.
This command will raise an execution error if the active trace in the selected channel is
not an eye diagram.
Suffix:
<Chn>
.
Channel number used to identify the active trace
Parameters:
<VoltageLevel>
Default unit: V
Options:
R&S
ZNB/ZNBT-K20
Manual operation:
See
CALCulate<Chn>:EYE:INPut:RTIMe:DATA
<RiseTime>
Sets/gets the rise time of the low pass in the binary signal generator simulation of the
related eye diagram.
This command will raise an execution error if the active trace in the selected channel is
not an eye diagram.
CALCulate<Chn>:EYE:STIMulus:LOWPass
. The
rise time definition can be modified using
CALCulate<Chn>:EYE:INPut:RTIMe:
.
Suffix:
<Chn>
.
Channel number used to identify the active trace
Parameters:
<RiseTime>
Default unit: s
Options:
R&S
ZNB/ZNBT-K20
Manual operation:
See
"Rise Time / Rise Time Definition"
CALCulate<Chn>:EYE:INPut:RTIMe:THReshold
<RiseThreshold>
Selects the appropriate rise time definition for the low pass in the binary signal genera-
tor simulation of the related eye diagram.
The rise time itself can be configured using
CALCulate<Chn>:EYE:INPut:RTIMe:
.
This command will raise an execution error if the active trace in the selected channel is
not an eye diagram.
Suffix:
<Chn>
.
Channel number used to identify the active trace
SCPI command reference