Command reference
R&S
®
ZNB/ZNBT
1069
User Manual 1173.9163.02 ─ 62
Query parameters:
<ChannelName>
Channel name, e.g.
'Channel 4'
.
Example:
See
Usage:
Query only
Manual operation:
See
CONFigure:CHANnel<Ch>[:STATe]
<Boolean>
Creates channel no.
<Ch>
and selects it as the active channel, or deletes it.
defines the channel name.
A channel created using
CONFigure:CHANnel<Ch>[:STATe] ON
can be configured
but has no trace assigned so that no measurement can be initiated. Use
CALCulate<Ch>:PARameter:SDEFine
<TraceName>, <Result>
to create a
channel and a trace. In remote control, it is possible to remove all channels. This is in
contrast to manual control where at least one channel with one diagram area and one
trace must be available.
Suffix:
<Ch>
.
Number of the channel to be created or deleted.
Parameters:
<Boolean>
ON - create channel no.
<Ch>
. If the channel no.
<Ch>
exists
already, it is not modified but selected as the active channel.
OFF - delete channel no.
<Ch>
.
*RST:
ON for channel no. 1 (created on *RST), OFF for all
other channels.
Example:
See
Manual operation:
See
CONFigure:CHANnel<Ch>:TRACe:CATalog?
Returns the numbers and names of all traces in channel no.
<Ch>
. The response is a
string containing a comma-separated list of trace numbers and names; see example. If
all traces have been deleted the response is an empty string ("").
Tip:
to query the traces in all channels of the
active recall set.
Suffix:
<Ch>
.
Channel number
Example:
See
Usage:
Query only
Manual operation:
See
SCPI command reference