Command reference
R&S
®
ZNB/ZNBT
1048
User Manual 1173.9163.02 ─ 62
Suffix:
<Ch>
.
Channel number
<PhyPt>
Physical port number
Setting parameters:
<Interchange>
FPORts | IPORts | SGATes
FPORts (or omitted)
Standard port sequence (network port 1 towards VNA, network
port 2 towards DUT)
IPORts | SGATes
Inverted port sequence (network port 2 towards VNA, network
port 1 towards DUT)
<arg1>
<block_data>
Content of a Touchstone file (
*.s2p
) in IEEE488.2
.
Usage:
Setting only
CALCulate<Ch>:TRANsform:VNETworks:SENDed:DEEMbedding<PhyPt>:
PARameters:G<Cmp>
<CircuitModel>, <Conductance>
CALCulate<Ch>:TRANsform:VNETworks:SENDed:DEEMbedding<PhyPt>:
PARameters:G<Cmp>?
<CircuitModel>
Specifies the conductance value G in the "shunt L, shunt C" circuit model for single
ended port deembedding.
In the query form, the <Conductance> parameter must be omitted.
Suffix:
<Ch>
.
Channel number
<PhyPt>
Physical port number
<Cmp>
Number of the conductance component in the circuit model. The
total number of conductances depends on the selected circuit
model.
Parameters:
<Conductance>
Conductance G<Cmp> for the specified circuit model.
Range:
-1kS to 1 kS.
Increment: 1 nS (1E-9 S)
*RST:
0 S
Default unit: Siemens (SI unit symbol: S)
Parameters for setting and query:
<CircuitModel>
SHLC
Circuit model whose conductance G<Cmp> shall be set, see
SCPI command reference