GUI reference
R&S
®
ZNB/ZNBT
388
User Manual 1173.9163.02 ─ 62
The appearance of the limit fail symbols is defined in the "Define User Color Scheme"
dialog (see
Chapter 5.17.3.2, "Define User Color Scheme dialog"
can choose between various options:
●
Change the trace color between failed measurement points.
●
Show or hide the colored squares.
Note:
●
Circle limit check and display of limit circles are independent of each other:
–
The limit circles can be displayed, no matter if the circle limit check is enabled.
–
If "Limit Check" is enabled, the ripple limits are checked, no matter if they are
displayed.
●
If result evaluation is limited to a user-defined
on page 386 is only performed inside this display circle.
●
The circle limits can only be checked if the trace format is complex. While a carte-
sian format is active, the limit circles are hidden and the circle limit check is sus-
pended.
●
If the limit check for (cartesian)
is enabled, it is also evaluated in complex
diagrams.
Figure 5-7: Simultaneous dB Mag limit line and circle check
Remote command:
CALCulate<Chn>:LIMit:CIRCle[:STATe]
CALCulate<Chn>:LIMit:CIRCle:FAIL?
CALCulate:LIMit:CIRCle:FAIL:ALL?
Clear Test
Resets the limit check results.
Remote command:
CALCulate<Chn>:LIMit:CIRCle:CLEar
Draw Circle
Activates touchscreen or mouse operation; tap the diagram at one border of the limit
circle and draw the circle to the required size and position.
Remote command:
n/a
Lines softtool