GUI reference
R&S
®
ZNB/ZNBT
312
User Manual 1173.9163.02 ─ 62
DUT
Balanced
port
Balanced
port
Log.
NWA
port
Physical port c
Logical ports j,k
(single-ended)
Physical port d
Logical port i
(balanced)
Physical port a
Physical port b
Imbalance 1 / Imbalance 2
Imbalance 1
selects the balanced logical port i as the input, the single-ended logical
ports j and k as the output and calculates the following imbalance parameter:
Imb
jk-i
= –(S
ca
– S
da
)/(S
cb
– S
db
)
Imbalance 2
selects the balanced logical port i as the output, the single-ended logical
ports j and k as the input and calculates the following imbalance parameter:
Imb
i-jk
= –(S
ac
– S
ad
)/(S
bc
– S
bd
)
Remote command:
CALCulate<Ch>:PARameter:MEASure "<Trace_Name>", "IMB<i>-<j><k>"
| "IMB<j><k>-<i>"
CALCulate<Ch>:PARameter:SDEFine "<Trace_Name>", "IMB<i>-<j><k>"
| "IMB<j><k>-<i>"
5.2.9
Stability tab
Selects one of the three two port stability factors K, μ
1
or μ
2
as measured quantities. A
typical application of stability factors is to assess the stability of an amplifier. Stability
factors cannot be calculated in balanced port configurations.