Command reference
R&S
®
ZNB/ZNBT
1252
User Manual 1173.9163.02 ─ 62
TYPE
Calibration type (see <Type> parameter in
]CORRection:COLLect[:ACQuire]:SELected
PORTs
The related test port numbers (comma-separated list of integers)
SPORt
For SMARTer calibrations this returns the port to which a power-
meter was connected (0 otherwise)
THRoughs
Measued Throughs: comma-separated list of ordered port num-
ber pairs <pno1-pno2>
TSTamp
Timestamp (in UTC)
LTSTamp
Timestamp in local time
TVNA
For calibrations involving switch matrices, this indicates the test
ports on the VNA itself. Returns a comma-separated list of port
pairs
TestPort,VnaPort
MVNA
For calibrations involving switch matrices, this indicates the
mapping between matrix ports and VNA ports for matrix
<Index>. Returns a comma-separated list of port pairs
MatrixVnaPort,VnaPort
MTESt
For calibrations involving switch matrices, this indicates the
mapping between matrix (physical) test ports and test ports for
matrix <Index>. Returns a comma-separated list of port pairs
MatrixTestPort,TestPort
CKIT
If created with FW version 2.30 or higher, for each port the name
of the used calibration kit is stored with the calibration.
The query can return one of the following:
–
the cal kit name
(string)
if available and unique
–
'Multiple'
if cal kit names are available but not unique, i.e. if <PhyPt> was
not specified and multiple cal kits were used (multi-port cals with
different connector types, merged cals)
'Unknown'
if cal kit names are not available (cals created with a FW version
< 2.30)
SCPI command reference