Command reference
R&S
®
ZNB/ZNBT
1367
User Manual 1173.9163.02 ─ 62
Suffix:
<Ch>
.
Calibrated channel number
Parameters:
<Boolean>
ON - other sources not necessarily switched off.
OFF - other sources switched off during the calibration sweep.
*RST:
ON
Example:
See
SOURce<Ch>:POWer:CORRection[:ACQuire]
Manual operation:
See
SOURce:POWer:CORRection:PMETer:ID
<PowerMeter>
Selects an external power meter for the scalar source power calibration (flatness cali-
bration).
Note:
The command cannot be used unless a power meter is connected via GPIB bus,
USB or LAN interface and configured in the "External Power Meters" dialog.
Parameters:
<PowerMeter>
Number of external power meter. The parameters UP, DOWN,
MIN, MAX are not available for this command.
Range:
1 to number of configured external power meters
*RST:
The power meter selection is not changed by a
reset of the analyzer.
Example:
See
SOURce<Ch>:POWer<PhyPt>:CORRection:COLLect[:
Manual operation:
See
SOURce:POWer:CORRection:PPOWer
<RecCalPower>
Defines the source power the R&S
ZNB/ZNBT uses to perform the first calibration
sweep of the source power calibration ("Reference Receiver Cal Power"). This prede-
fined power value is ignored if the R&S
ZNB/ZNBT is set to use the calculated port
power result (
SOURce:POWer:CORRection:PSELect
CPOW
).
Parameters:
<RecCalPower>
Reference receiver cal power
Range:
Depending on source power range of the analyzer
and the power range of the external power meter.
*RST:
0 dBm
Default unit: dBm
Example:
See
SOURce<Ch>:POWer:CORRection[:ACQuire]
Manual operation:
See
"Reference Receiver Cal Power"
SCPI command reference