GUI reference
R&S
®
ZNB/ZNBT
315
User Manual 1173.9163.02 ─ 62
5.2.11.1
Controls on the DC tab
The "PAE..." button opens the
dialog.
DC 1/DC 2/DC 3/DC 4/Source Port
Selects a DC INPUT and the drive port for the active trace.
The DC voltages "DC 1", ..., "DC 4" are measured at the rear panel connectors DC
Input 1, ..., 4, respectively. For best accuracy, it is recommended to adjust the
"Ranges" to the measured signal.
Remote command:
CALCulate<Ch>:PARameter:MEASure
"<Trace_Name>", "DC1d1" |
"DC2d1" | ...
CALCulate<Ch>:PARameter:SDEFine
"<Trace_Name>", "DC1d1" |
"DC2d1" | ...
Ranges
"DC 1", ..., "DC 4" in the "Ranges" section configure the measurement ranges of DC
Input 1, ..., 4, respectively. For best accuracy, adjust the ranges to the measured sig-
nals. To protect the instrument, all ranges are pre-set to "+/- 20 V" (the widest range).
If the actual voltage of the input signal exceeds the selected range, the measurement
is incorrect, and a DC overload message is displayed.
Remote command:
5.2.11.2
Power Added Efficiency (PAE) dialog
The "Power Added Efficiency (PAE)" dialog configures the measurement of the PAE of
an active 2-port device.
Access:
Trace – [Meas] > "DC" > "PAE..."
Meas softtool