Command reference
R&S
®
ZNB/ZNBT
1240
User Manual 1173.9163.02 ─ 62
[SENSe:]CORRection:COLLect:PMETer:ID
<PowerMeter>, <TestPort>
Selects an external power meter for the SMARTerCal and assigns it to an analyzer
port.
Note:
The command cannot be used unless a power meter is connected via GPIB bus,
USB or LAN interface and configured in the "External Power Meters" dialog.
Parameters:
<PowerMeter>
Number of external power meter. The parameters UP, DOWN,
MIN, MAX are not available for this command.
Range:
1 to number of configured external generators
*RST:
The power meter selection is not changed by a
reset of the analyzer.
<TestPort>
Test port number. During power calibration, the power meter is
connected to this port.
*RST:
1 to the number of test ports
Example:
See
[SENSe<Ch>:]CORRection:COLLect:AUTO:POWer
Manual operation:
See
[SENSe<Ch>:]CORRection:COLLect:SAVE:SELected:DEFault
Generates a set of default system error correction data for the selected ports and cali-
bration type. The default data set corresponds to a test setup which does not introduce
any systematic errors; none of the measurement results acquired previously
(
[SENSe<Ch>:]CORRection:COLLect[:ACQuire]:SELected
) is taken into
account.
Tip:
The main purpose of the default correction data set is to provide a dummy system
error correction which you can replace with your own, external correction data. You
may have acquired the external data in a previous session or even on an other instru-
ment. If you want to use the external correction data on the analyzer, simply generate
the default data set corresponding to your port configuration and calibration type and
overwrite the default data. For details refer to the program example below.
Suffix:
<Ch>
.
Channel number of the calibrated channel.
SCPI command reference