GUI reference
R&S
®
ZNB/ZNBT
314
User Manual 1173.9163.02 ─ 62
Power Meter
Shows a list of all power meters that have been properly configured. See
The bordered label below displays the type and serial number of the selected power
meter.
Remote command:
CALCulate<Ch>:PARameter:MEASure
'TraceName', 'PmtrD1 | ...
CALCulate<Ch>:PARameter:SDEFine
'TraceName', 'PmtrD1 | ...
Auto Zero
Initiates an automatic zeroing procedure of the selected power meter.
The power meter must be disconnected from the RF power; see
on page 266. A message indicates that zeroing is finished.
Remote command:
SYSTem:COMMunicate:RDEVice:PMETer<Pmtr>:AZERo
Source Port
Selects one of the available test ports of the analyzer as a source of the stimulus sig-
nal.
Remote command:
CALCulate<Ch>:PARameter:MEASure
'TraceName', 'PmtrD1 | ...
CALCulate<Ch>:PARameter:SDEFine
'TraceName', 'PmtrD1 | ...
Show as
Selects the physical unit of the displayed trace. It is possible to display the measured
"Voltage" V or convert it to a "Power" according to the formula
P = V
2
/Re(Z
0
).
Z
0
denotes the reference impedance of the source port. The reference impedances are
defined in the "Balanced Ports" dialog (see
Chapter 5.2.1.3, "Balanced Ports dialog"
Remote command:
5.2.11
DC tab
Selects the ranges for the DC voltages fed to the DC Input connectors at the rear panel
as measured quantities and opens the PAE measurement dialog. The DC input con-
nectors are available as option R&S
ZNB/ZNBT-B81 (see
Meas softtool