[SENSe<Ch>:]FREQuency:CONVersion:MIXer:LOMultiplier<Stg>
<Numerator>,
<Denominator>
Selects the frequency conversion factors for the LO signal.
The conversion factor is a ratio of two positive integers: numerator and denominator.
Suffix:
<Ch>
.
Channel number
<Stg>
Mixer stage signals LO 1 or LO 2; see
]FREQuency:CONVersion:MIXer:STAGes
Parameters:
<Numerator>
Range:
1, 2, 3 ...
*RST:
1
<Denominator>
Range:
1, 2, 3 ...
*RST:
1
Example:
See
[SENSe<Ch>:]FREQuency:CONVersion:MIXer:
Manual operation:
See
[SENSe<Ch>:]FREQuency:CONVersion:MIXer:LOPort<Stg>
<SourceType>[,
<SourceNumber>]
Selects an analyzer or an external generator port as a signal source for the LO1 or
LO2 signal.
Note:
If
are part of the RF connection configuration, operation with
more than one internal source is
not
supported. In this case only external generators
can be used as local oscillators.
Suffix:
<Ch>
.
Channel number
<Stg>
Mixer stage (signals LO 1 or LO 2; see
]FREQuency:CONVersion:MIXer:STAGes
Parameters:
<SourceType>
PORT | GENerator | NONE
NONE
–
LO signal is not controlled by the analyzer.
PORT
–
analyzer port
GENerator
–
external generator
*RST:
NONE
<SourceNumber>
Number of analyzer port or generator.
Example:
See
[SENSe<Ch>:]FREQuency:CONVersion:MIXer:
Manual operation:
See
SCPI command reference