Wide pedal: You can temporarily activate the manual chip blow by pushing
and holding the pedal to the right.
The flow of air will continue until you remove your foot from the pedal.
NOTE
The type of manual chip blow does not depend on the type of the automatic
chip blow, but is always dry. The manual chip blow can be set to be
dependent on programming; contact your Planmeca dealer.
20.4.7 Instrument light
The instrument light is either on or off when you pick up the instrument from
the instrument console.
Whether the light is on or off, as well as its intensity, can be defined in the
programming mode, see section "Instrument light" on page 100.
20.5 Scaler
CAUTION
Do not use the scaler on patients with cardiac pacemakers. The scaler can
cause disturbance on the pacemaker’s function.
NOTE
A scaler requires additional electronics and the scaler type can not be
changed without changing electronics.
20.5.1 Speed/power
Standard pedal: To drive the instrument, push the foot control pedal to the
right or to the left.
Wide pedal: To drive the instrument, push the foot control pedal down.
20 Operating instruments
84 Planmeca Compact i
User's manual