24.5.3.1 OroCup
About this task
The following describes how to disinfect the suction system with OroCup.
Steps
1. Mix the cleaning solution:
Pour 20 ml Planmeca approved suction disinfectant into the rinsing
bottle. Add 1 l water and shake well.
NOTE
Do not use dish washing detergents.
2. Place the rinsing bottle (for example OroCup) on a flat surface (table or
floor).
3. Remove the suction handpieces from their holders and push them onto
the inserts inside the cap of the rinsing bottle. Allow the bottle to empty.
4. Return the suction tubes back to the holder right after OroCup has been
emptied. Do not use suction only for air.
5. Replace the exchangeable parts (filters etc.) the next morning.
24.5.4 Weekly cleaning procedures
24.5.4.1 During working day
About this task
Clean the suction system with Dürr MD 555 cleaner 1 - 2 times a week to
prevent the buildup of deposits in the suction system, especially if you are
using dental air polishers.
This procedure is mandatory for dental units with a Dürr amalgam separator
or Dürr VS/A separator.
For more information on Dürr MD 555 cleaner, please visit
.
Steps
1. Pour 50 ml of Dürr MD 555 cleaner into a rinsing bottle (for example
OroCup). Add 1 l water and mix well.
24 Cleaning and disinfection
User's manual
Planmeca Compact i 125
Summary of Contents for Compact i Classic v2
Page 1: ...PlanmecaCompact iClassicv2 user s manual EN 10032684...
Page 8: ...Table of contents Planmeca Compact i User s manual...
Page 23: ...8 Planmeca Compact i dental unit User s manual Planmeca Compact i 15...
Page 114: ...4 Press Program 22 Programming 106 Planmeca Compact i User s manual...
Page 185: ...30 Technical information User s manual Planmeca Compact i 177...
Page 189: ......