Help messages
HELP
CODE
MEANING
ACTION OR EXPLANATION
H 8
Chair height can not be programmed this
high in automatic positions for patient safety
reasons.
Run the chair slightly down to be able to
program this position into memory.
H 9
This scaler cannot currently be used with
this unit.
This scaler needs a dedicated electronics
that is not installed, or the scaler electronics
installed is not compatible with this scaler.
H 10
Wait while configuring the dental instrument
setup.
Wait a few seconds for the unit to check
your instrument configuration, since it has
changed.
H 11
Software download mode is on.
H 12
Unit's internal error reporting is disabled.
The unit operates normally but no errors are
reported (exhibition use). Use service mode
n.15 to turn error reporting back on.
H 13
Instrument cannot run because of patient
security violation.
Return the foot control pedal once to the
middle position to start the instrument.
Instruments do not start if the standard foot
control pedal is already pushed to the right
or left when the instrument is picked up from
the instrument console.
H 14
This dental instrument is not recognised
(new type of instrument).
To be able to operate this new instrument
the MCB software must be updated.
Contact your Planmeca dealer.
H 14.1 Instrument presets reset to defaults.
Instrument presets are faulty. Presets are
reset to defaults.
H 15
This button has no function at the moment.
Instruments: Pick up a dental instrument
before attempting to alter its settings; or if
the instrument is selected, this function is
not allowed with this instrument. Other
buttons: the unit does not have this function.
H 16
Operating light is dimmed and
polymerisation light can be used.
This feature enables easier work with
composite materials, minimising the risk of
operating light induced curing. This feature
is triggered by picking up and returning the
polymerisation light to its holder without
starting it. The operating light returns to
normal at the same instance any control is
touched or another instrument selected. The
feature can be disabled.
Contact your Planmeca dealer.
H 17
No syringe connected to instrument
console.
The console syringe must be re-connected,
otherwise there is an air-leak from the
syringe QC whenever an instrument or
device that requires air pressure is used.
H 18
No instrument hose connected to this
instrument place.
The instrument is picked up, but there is no
hose connected at this location. Check that
all hoses go to the right instrument arms.
27 Help and error messages
142 Planmeca Compact i
User's manual
Summary of Contents for Compact i Classic v2
Page 1: ...PlanmecaCompact iClassicv2 user s manual EN 10032684...
Page 8: ...Table of contents Planmeca Compact i User s manual...
Page 23: ...8 Planmeca Compact i dental unit User s manual Planmeca Compact i 15...
Page 114: ...4 Press Program 22 Programming 106 Planmeca Compact i User s manual...
Page 185: ...30 Technical information User s manual Planmeca Compact i 177...
Page 189: ......