1. Turn the bowl and the cup fill tube away from above the cuspidor.
2. Remove the bowl filter parts.
3. Push the bowl fastening clip carefully down and
4. rotate the bowl counter-clockwise.
5. Lift the bowl slightly upward and remove it by pulling it horizontally.
Replace the bowl in reverse order.
NOTE
The cover cap of the filter makes the removal of the filter easier, but the filter
can also be used without the cover cap.
24.4.2 Instrument flushing holder
The instrument flushing holder can be lifted away from its position.
Disinfect the flushing holder once a month in a washer-disinfector at 93°C or
autoclave it at 134°C.
24 Cleaning and disinfection
122 Planmeca Compact i
User's manual