Help messages
HELP
CODE
MEANING
ACTION OR EXPLANATION
H 30
Close water tap. Fill container with
disinfectant up to groove and attach to unit.
H 31
Switch unit off, leave disinfectant in unit
overnight (min. 8 h).
When unit is turned back on, disinfection
procedure continues.
H31.1
Remove container, fill it up to groove with
water and attach to unit.
H31.2
Attach container filled with water.
H 32
Container flushing completed.
Remove container from unit.
H 33
Open water tap.
H 34
No instruments selected when starting
instrument flushing or waterline cleaning.
Select at least one instrument and syringe
(remove from holder, bend the instrument
arms).
H 35
Instruments are locked for safety reasons.
Select instruments, place them in flushing
holder, and press Instrument flushing button
to start instrument flushing.
H35.1
Waterline cleaning cancelled or failed.
Remove container.
Make sure water tap is open. Select
instruments, place them in flushing holder
and press Instrument flushing button to start
instrument flushing.
H 36
Flushing ready. Return instruments to
instrument console.
Also bend and release syringe's instrument
arm once.
H36.1
Instrument flushing completed, but water
flow was not detected for one or more
instruments.
Run instruments for a while with spray water
to ensure instrument flushing is sufficient.
H36.8
Low water pressure during final flushing.
Perform long instrument flushing to ensure
instruments are flushed.
H 37
Waterline cleaning has been interrupted.
H 38
Low water pressure during final rinsing.
When waterline cleaning is finished, perform
long instrument flushing to ensure
instruments are flushed.
H 39
Syringe misplaced or its flow not detected in
waterline cleaning.
Return syringe to console. Then, select
syringe (bend instrument arm) again and
place it carefully in flushing holder.
H 40
Too many Bien-Air MX brushless
micromotors.
A maximum of two is allowed.
H 41
Insert suction tubes to suction tube cleaning
holder, add disinfectant and press green
start button (upper button).
H 41.1 Insert suction tubes to suction tube cleaning
holder.
H 41.2 Remove suction tubes from holder to start
suction cleaning.
H 42
Suction cleaning completed. Return suction
tubes to suction holder and close suction
cleaning cover.
27 Help and error messages
144 Planmeca Compact i
User's manual
Summary of Contents for Compact i Classic v2
Page 1: ...PlanmecaCompact iClassicv2 user s manual EN 10032684...
Page 8: ...Table of contents Planmeca Compact i User s manual...
Page 23: ...8 Planmeca Compact i dental unit User s manual Planmeca Compact i 15...
Page 114: ...4 Press Program 22 Programming 106 Planmeca Compact i User s manual...
Page 185: ...30 Technical information User s manual Planmeca Compact i 177...
Page 189: ......