22 Programming
22.1 Introduction
NOTE
If the instrument is operated during programming, the changes in settings
are seen immediately.
NOTE
The air/water flow of the syringe can not be programmed.
NOTE
To check a function setting without programming it, press Program and the
button of the desired function. The setting is displayed on the control panel.
Press Program again to close the window without changing the setting.
Most of the programming follows the same pattern:
1. If you are programming an instrument, pick up the instrument from the
instrument console.
2. Press Program to activate the programming mode.
The indicator light is lit.
3. Select the desired function from the control panel.
4. Change the setting of the function with the foot control. To increase the
parameter value, push the pedal to the right. To decrease the parameter
value, push the pedal to the left. When the pedal is pushed halfway (right
or left) the parameter changes slowly. When the pedal is pushed to its
extreme position (right or left) the parameter changes quickly. If the
pedal is pushed for longer than 0.4 seconds, the value will change as
long as the pedal is pushed.
NOTE
Alternatively, you can change the settings with the up/down buttons. The
instrument settings can be adjusted with these buttons without operating
the instrument.
5. Press Program to store the new setting into memory.
NOTE
If you interrupt programming for over 90 seconds, the unit automatically exits
the programming mode without storing the new settings.
22.2 Automatic chair positions
NOTE
The chair height cannot be programmed near the upper limit. The help code
H 8 is displayed if the chair position is not allowed. When needed, the upper
limit can be altered. Contact your Planmeca dealer.
22 Programming
User's manual
Planmeca Compact i 97
Summary of Contents for Compact i Classic v2
Page 1: ...PlanmecaCompact iClassicv2 user s manual EN 10032684...
Page 8: ...Table of contents Planmeca Compact i User s manual...
Page 23: ...8 Planmeca Compact i dental unit User s manual Planmeca Compact i 15...
Page 114: ...4 Press Program 22 Programming 106 Planmeca Compact i User s manual...
Page 185: ...30 Technical information User s manual Planmeca Compact i 177...
Page 189: ......