Prerequisites
•
The intraoral camera must be connected to the USB port.
When the intraoral camera is placed in the instrument console, it is
connected to the USB port on the underside of the instrument console.
When the intraoral camera is placed in the suction holder, it is connected
to the USB port on the cuspidor.
•
Planmeca Romexis software must be installed and the connection
between it and the dental unit must be enabled. When Planmeca
Romexis is running, the intraoral camera is continuously connected to
the software.
After you have saved the images
Once you have saved the images, you can browse through them in the
Planmeca Romexis Image browser. Press the Done button at the bottom of
the intraoral camera window in Planmeca Romexis to move to the Image
browser. For more information on the Image browser, see the
Planmeca
Romexis user’s manual.
20.7.1 Intraoral camera placed in instrument console
Starting intraoral camera
Before you turn on the intraoral camera, first select the patient and then
select the 2D module in Planmeca Romexis.
To turn on the intraoral camera, pick it up from the instrument console.
Alternatively, you can turn the intraoral camera on by pressing either of the
buttons on the camera handpiece (Power/macro or Image control button).
NOTE
If you turn on the intraoral camera from the buttons on the handpiece, the
camera can not be operated from the dental unit's foot control.
When the intraoral camera turns on,
•
the indicator light on the handpiece turns blue,
•
you will hear a short sound signal,
•
the Planmeca Romexis intraoral camera view is displayed on the
monitor,
•
the text “I_CAM” shows on the display.
Once the intraoral camera is activated, you can operate it
•
from the dental unit's foot control, if the camera has been turned on by
picking it up from the instrument console
•
from the camera handpiece, or
•
remotely from the Planmeca Romexis user interface.
You can use the camera in normal or macro mode, as well as freeze and
save the image. The LED on the handpiece is turned on and off
simultaneously with the intraoral camera.
Macro on/off
A short sound signal indicates that the macro mode is turned on/off. When
the macro mode is on, autofocus is off and the indicator light on the
handpiece is lilac.
20 Operating instruments
User's manual
Planmeca Compact i 89
Summary of Contents for Compact i Classic v2
Page 1: ...PlanmecaCompact iClassicv2 user s manual EN 10032684...
Page 8: ...Table of contents Planmeca Compact i User s manual...
Page 23: ...8 Planmeca Compact i dental unit User s manual Planmeca Compact i 15...
Page 114: ...4 Press Program 22 Programming 106 Planmeca Compact i User s manual...
Page 185: ...30 Technical information User s manual Planmeca Compact i 177...
Page 189: ......