Help messages
HELP
CODE
MEANING
ACTION OR EXPLANATION
H 19
No syringe connected to connector located
on cuspidor.
The assistant's syringe must be re-
connected, otherwise there is an air-leak
from the assistant's syringe QC whenever
an instrument or device that requires air
pressure is used.
Contact your Planmeca dealer.
H 20
Wrong instrument is connected to cuspidor
syringe connector or assistant instrument
configuration is wrong.
Contact your Planmeca dealer.
H 21
Flow of both spray air and spray water are
programmed to zero (no flow).
Contact your Planmeca dealer.
H 22
Main air or water valve is set to be always
closed in SM 12 (air) /13 (water).
This is a reminder that either of the main
valves has been set to be always closed.
The setting may be changed in service
modes SM 12 (air) and SM 13 (water).
H 23
Automatic/manual chip blow cannot be used
when sterile water function is selected.
H 24
Programming of selected setting is disabled. The programming of the selected setting is
disabled.
Contact your Planmeca dealer.
H 25
Planmeca Compact e help message:
Patient chair cannot be driven into
automatic chair position.
Contact your Planmeca dealer.
H 26
Drive chair upwards, then lower backrest. If
patient chair is attached to unit, chair's
automatic positions are not switched on.
Drive the chair upwards before lowering the
backrest. If this help message is shown
while the patient chair is attached to the
unit, it means that you have forgotten to
switch on the chair’s automatic positions. To
clear the message, put the operation mode
switch on the cuspidor side of the chair base
in the “AUTO” position (switch up).
H 27
Drive backrest upwards, then lower chair. If
patient chair is attached to unit, chair's
automatic positions are not switched on.
Drive the backrest upwards before lowering
the chair. If this help message is shown
while the patient chair is attached to the
unit, it means that you have forgotten to
switch on the chair’s automatic positions. To
clear the message, put the operation mode
switch on the cuspidor side of the chair base
in the “AUTO” position (switch up).
H 28
Planmeca Compact WE's help message.
Connect the lifter cable.
H 29
Drive chair downwards, then lower backrest.
If patient chair is attached to unit, chair's
automatic positions are not switched on.
Drive the chair downwards before lowering
the backrest. If this help message is shown
while the patient chair is attached to the
unit, it means that you have forgotten to
switch on the chair’s automatic positions. To
clear the message, put the operation mode
switch on the cuspidor side of the chair base
in the “AUTO” position (switch up).
27 Help and error messages
User's manual
Planmeca Compact i 143
Summary of Contents for Compact i Classic v2
Page 1: ...PlanmecaCompact iClassicv2 user s manual EN 10032684...
Page 8: ...Table of contents Planmeca Compact i User s manual...
Page 23: ...8 Planmeca Compact i dental unit User s manual Planmeca Compact i 15...
Page 114: ...4 Press Program 22 Programming 106 Planmeca Compact i User s manual...
Page 185: ...30 Technical information User s manual Planmeca Compact i 177...
Page 189: ......