Power supply related error messages
ERROR
CODE
EXPLANATION
E11.6
Water heater fuse (F6) has blown on the MCB.
E11.7
Syringe heater fuse (F7) has blown on the MCB.
E11.8
Polymerisation light fuse (F8) has blown on the MCB.
E12.1
SELV voltage too low. Unit functionality may be limited.
E12.2
SELV voltage too high.
E13
Mains frequency outside acceptable range.
E14
IPS temperature measurement sensor error.
E15
IPS heatsink is running too hot on MCB.
E16
Attached scaler electronics is of new type.
E16.1
New type of scaler electronics is attached to MCB.
Update MCB software.
E16.2
New type of scaler electronics is attached to IMUX.
Update MCB software.
E16.3
Scaler electronics slot does not match with scaler hose
connection.
Check scaler electronics connection.
E16.4
Other instrument than scaler is connected to slot where scaler
electronics is.
Check scaler electronics connection.
E17
Power Supply related error messages.
E18
Power Supply related error messages.
27.3.3 Control panel related error messages
Control panel related error messages
ERROR
CODE
EXPLANATION
E19.1
Control panel in branch 1 is of new type.
Update MCB software.
E19.2
Control panel in branch 2 is of new type.
Update MCB software.
E20.1
Control panel 1 button stuck.
Control panel button is stuck during self test.
E20.2
Control panel 2 button stuck.
Control panel button is stuck during self test.
E21
Control panel related error messages.
E22
Control panel related error messages.
27 Help and error messages
User's manual
Planmeca Compact i 153
Summary of Contents for Compact i Classic v2
Page 1: ...PlanmecaCompact iClassicv2 user s manual EN 10032684...
Page 8: ...Table of contents Planmeca Compact i User s manual...
Page 23: ...8 Planmeca Compact i dental unit User s manual Planmeca Compact i 15...
Page 114: ...4 Press Program 22 Programming 106 Planmeca Compact i User s manual...
Page 185: ...30 Technical information User s manual Planmeca Compact i 177...
Page 189: ......