Saliva suction handpiece
The saliva suction handpiece can be disassembled for cleaning by pulling
out the end of the handpiece and removing the core (2) and the adjuster (1).
The handpiece parts can be disinfected in a washer-disinfector at 93°C, then
optionally autoclaved at 134°C.
24.6 Planmeca WEK water disinfection system
For information on how to use the Planmeca WEK water disinfection system,
see
Planmeca WEK water disinfection system user's manual.
24.7 Planmeca intraoral scanner
For information on how to clean the Planmeca intraoral scanner, see
Planmeca FIT user’s manual.
24.8 External PC
The external PC together with its mouse and keyboard can be wiped with a
dry cloth, or according to the manufacturer’s instructions.
CAUTION
When cleaning the external PC, always disconnect the PC from the mains
electricity supply.
24 Cleaning and disinfection
128 Planmeca Compact i
User's manual