22.5.3 Maximum brightness
Steps
1. Press Program.
2. Press Operating light.
3. Press chair position button A.
The letter I and 1-3 horizontal lines appear on the display.
The lines indicate the maximum brightness of the operating light.
•
One line = mild
•
Two lines = strong
•
Three lines = glaring
4. Change the maximum brightness by pressing the chair position button A
for one second. Every time you press the button, the maximum
brightness changes. The number of horizontal lines changes
accordingly.
5. Press Program.
22.6 Duration of door open / assistant call
Steps
1. Press Program.
2. Press Door open / assistant call.
3. Adjust the door open / assistant call time.
The door open / assistant call time is displayed on the control panel.
The minimum value is 0 seconds, the maximum 250 seconds. The value
is adjusted in steps of 1 in value range 0 - 30, and in steps of 5 in value
range 30 - 250. The default value is 5.
22 Programming
User's manual
Planmeca Compact i 105