Help messages
HELP
CODE
MEANING
ACTION OR EXPLANATION
H 81
Headrest angle B safety feature activated
when driving headrest. Check that nothing is
blocking headrest movements.
Check that nothing is blocking the headrest
movement. The headrest can be driven
normally after the possible obstruction has
been removed.
Faulty angle B motor, motor cable or
position sensor may cause this problem.
H81.1
Headrest angle A drive limit activated when
driving headrest.
Check that nothing is blocking the headrest
movement.
Faulty angle A motor, motor cable or
position sensor may cause this problem.
H 82
Connecting to wireless foot control was
successful.
Connection was established with the
requested foot control.
H 83
Foot control handle pressed down.
Release the foot control handle.
H 84
Connecting to wireless foot control failed.
Connection with the requested foot control
could not be established. Make sure the foot
control is active (connect to a charger, if
possible), is in the vicinity of the unit, and try
again.
H 85
Software update denied. Chair was moving
or instrument was in use while starting
software update.
The chair was moving or an instrument was
in use while starting the software update.
H 86
Software update denied. Incorrect control
panel type was detected while starting
software update.
An incorrect control panel type was detected
while starting the software update.
H 87
Software update halted. Activate the
wireless foot control by pressing the handle.
Wireless foot control has to be activated
before it can be updated.
H 88
Wireless foot control configuration was
successful.
Wireless foot control channel and power
settings were successfully changed.
H 89
Wireless foot control configuration failed.
Wireless foot control channel and power
settings could not be changed. Make sure
the foot control is active (connect to a
charger, if possible), is in the vicinity of the
unit, and try again.
H 90
This button has no function at the moment,
press Pos Sel, Chair/Headr/Light Sel first.
H 90.1 Measure water chlorine level.
Run at least 50 ml of water from the syringe
to the cup. Move the test strip back and
forth in the water for 30 seconds. Compare
the strip colour with the colour scheme.
Select the chlorine level on the control panel
according to the measurement.
< 0.2 (LOW)
0.2 - 0.8 (OK)
> 0.8 (HIGH)
If the chlorine level is continuously too low,
contact service.
27 Help and error messages
User's manual
Planmeca Compact i 147
Summary of Contents for Compact i Classic v2
Page 1: ...PlanmecaCompact iClassicv2 user s manual EN 10032684...
Page 8: ...Table of contents Planmeca Compact i User s manual...
Page 23: ...8 Planmeca Compact i dental unit User s manual Planmeca Compact i 15...
Page 114: ...4 Press Program 22 Programming 106 Planmeca Compact i User s manual...
Page 185: ...30 Technical information User s manual Planmeca Compact i 177...
Page 189: ......