Main CPU related error messages
ERROR
CODE
EXPLANATION
E79
IMUX water pressure cannot be released during waterline
cleaning cycle.
E79.1
IMUX water pressure rose during waterline cleaning cycle
when pressure rise was not expected.
27.3.12 Headrest related error messages
Headrest related error messages
ERROR
CODE
EXPLANATION
E80.1
Motorised headrest is not found (according to service mode
configuration).
E80.2
Motorised headrest is found, but is not set in service mode
configuration.
E80.3
Headrest joystick (dentist side) stuck or cable broken.
E80.4
Headrest joystick (assistant side) stuck or cable broken.
E80.5
Motorised headrest motor A potentiometer or cable is faulty or
disconnected.
E80.6
Motorised headrest motor B potentiometer or cable is faulty or
disconnected.
E80.7
Motorised headrest length potentiometer or cable is faulty or
disconnected.
E81
Motorised headrest communication error (may be too old or
wrong software or loose cables or short circuit).
E81.1
Motorised headrest communication error.
E82
Headrest software is incompatible with MCB software.
Headrest or MCB software should be upgraded.
E82.1
Headrest software is incompatible with MCB software.
Headrest or MCB software should be upgraded.
E82.2
Incompatible headrest software.
27.3.13 Water leak related error messages
Water leak related error messages
ERROR
CODE
EXPLANATION
E90
Water leak detected.
Main water valve is closed.
E90.1
Water leak detected. Water use allowed temporarily.
27 Help and error messages
160 Planmeca Compact i
User's manual
Summary of Contents for Compact i Classic v2
Page 1: ...PlanmecaCompact iClassicv2 user s manual EN 10032684...
Page 8: ...Table of contents Planmeca Compact i User s manual...
Page 23: ...8 Planmeca Compact i dental unit User s manual Planmeca Compact i 15...
Page 114: ...4 Press Program 22 Programming 106 Planmeca Compact i User s manual...
Page 185: ...30 Technical information User s manual Planmeca Compact i 177...
Page 189: ......