24 .
RamanMicro 200 Series Getting Started Guide
Where
λ
is the laser wavelength in micrometers (
µ
m). For example, a laser wavelength of
785 nm (0.785
µ
m) will have an
MPE
value of 1.49 x 10
−2
J cm
−2
or 1.49 x 10
−3
W cm
−2
for
a 10 second exposure or longer. Table 3 shows the calculated nominal ocular hazard distance
for 785 nm Raman systems.
Table 3 Nominal Ocular Hazard Distance (NOHD) for 785 nm
Probe focal length / mm
NOHD for 100 mW exiting fiber optic probe / cm
7.5 (± 10%)
27.3
NOTE: Other focal lengths are available; contact PerkinElmer for details. In particular, the
use of an ultra-long working distance lens adaptor (part number L1320071) provides
a working distance of 25 mm. Where different focal lengths are used, the NOHD
should be calculated accordingly.
Internal Interlock System
There is no internal interlock system fitted to the RamanMicro 200. If instrument panels are
removed collimated laser light may be visible. Under no circumstances remove any
instrument panels.
External Interlock System
Class 3B laser systems have the potential to cause harm or injury to those using them or
others in the vicinity of the instrument. For more information refer to
Attaching the External
Interlock
on page 40.
Summary of Contents for RamanMicro 200 Series
Page 1: ...RamanMicro 200 Series Getting Started Guide MOLECULAR SPECTROSCOPY...
Page 5: ...Introduction...
Page 10: ...10 RamanMicro 200 Series Getting Started Guide...
Page 11: ...Safety Practices...
Page 27: ...Raman History Theory and Instrumentation...
Page 33: ...Preparing Your Facility...
Page 37: ...Installing Your RamanMicro 200 Series...
Page 57: ...Using the RamanMicro 200 Series...
Page 89: ...Maintenance and Troubleshooting...
Page 98: ...98 RamanMicro 200 Series Getting Started Guide...