Preparing Your Facility . 35
Finding a Suitable Location for the Instrument
The following conditions should be observed:
•
The environment must be suitable for a Class 3B laser system.
•
There must be a minimum of 4 power points (1 for the microscope, 1 for the
spectrometer and 2 for the PC). If your system has a motorized stage you will also
require 1 power point for the stage controller.
•
The workbench must be flat and level.
•
The workbench where the spectrometer and PC are to be located should be at a
suitable height to allow the user to work comfortably.
•
Use conducting floor coverings to minimize static discharge.
•
Floor vibrations or noise from heavy manufacturing equipment can affect the
performance of the instrument.
•
Intense magnetic fields should be at least 5.5 meters (18 feet) away from the
spectrometer.
•
Never place the instrument near a window that gets direct sunlight.
Ensure all users understand laser safety issues before locating and using the instrument.
Consult your Laser Safety Officer and perform a risk assessment before installing the
equipment.