Installing Your RamanMicro 200 Series . 53
10.
Ensure that you have followed the instructions on the screen, and then click
Next
.
The following dialog is displayed.
11.
Adjust the
Laser Power
until you can see a clear visible spot.
12.
If the laser spot is visible and close to the cross hairs, you do not need to physically
adjust the camera module. Continue at step 16 to perform a fine adjustment to align
the cross hairs with the laser spot.
OR
If no spot is visible, a physical adjustment of the camera module is required. Continue
at step 13.
13.
Loosen the 4 screws on the top plate of the Raman module until the camera is just free
to move (Figure 24).
Figure 24 Top view of the microscope showing the screws used to adjust the
position of the camera module
14.
Move the camera module until the crosshairs line up as closely as possible with the
center of the laser spot.
15.
Re-tighten the 4 screws.