40 .
RamanMicro 200 Series Getting Started Guide
Attaching the External Interlock Override
The RamanMicro 200 Series has an external interlock connection located on the rear panel of
the spectrometer module. To operate the instrument, the external interlock must be
satisfied.
You must connect an external door interlock to the 7-way DIN connector labeled
INTERLOCK on the rear panel of the spectrometer, or attach the External Interlock Override
(L1320264) supplied with the instrument.
If your laser safety officer determines that an interlock is required on the door to the room,
you will need to purchase the Raman Laser Safety Kit (L1320262). For more information
contact your PerkinElmer representative.