Installing Your RamanMicro 200 Series . 43
Attaching the Z-axis Focus Motor and Sleeve
NOTE: If you have purchased the optional microscope light-proof enclosure, attach the z-
axis focus motor and sleeve before fitting the enclosure. See the
Raman Microscope
Light-Proof Enclosure
leaflet (L1321873) for more information.
1.
Slide the focus motor sleeve over the z-axis drive control knob on the left side of the
microscope, and then tighten the connections using a hexagonal key.
2.
Push the focus motor into the sleeve until it touches the drive control, and then tighten
the screw to finger tight.
3.
Connect the serial connector from the focus motor into the connection at the rear of the
stage control box labeled MOTOR Z (Figure 15).
Figure 16 Z-axis focus motor and sleeve attached to the microscope inside the
optional light-proof enclosure
NOTE: Ensure that the pre-focusing lever on the coarse adjustment control is on the off
position.