Chapter 4 Description of Functions
4-103
4
Interpolation method
When the displacement data of the shape waveform acquired from the sensor
is the alarm value (+999.999), select the method to display it on the plot
screen after eliminating the alarm value.
* The data of abnormal value processing in the
"Common tab"
settings is
processed in the sensor. Interpolation in this
"System settings"
is shown
only on the display of the HL-D3SMI. It does not change the displacement
measurement value.
Settings
Illustration
Description
No interp.
Displays the acquired data without
interpolation.The alarm value
(+999.999) is displayed as is.
(
in the figure shows the original
displacement which resulted in the
alarm value due to insufficient
beam.)
Segm. interp.
Interpolate between normal dis-
placement measurement values after
removing alarm values in acquired
data.
Cut line
Erase data between measurement
values after removing alarm values.
Changes the
color of inter-
polated data.
Check this when "Segm. interp." is
selected. The interpolated line seg-
ment is displayed in different color
from the measurement value.