9
Head Impulse
9.1
Test environment
The environment can vary but must allow you to position the patient at least one meter from the wall (or other solid sur-
face that can be used as a projection surface).
For Calibration and Test Purposes:
Note
•
The patient should be seated in a chair that is stationary and does not swivel.
1. Choose a wall that allows you to position the patient at least one meter in front of the fixation dot.
2. Apply one of the fixation dots supplied with the system to the wall in a location that allows you to position the patient
directly in front of the fixation dot.
9.2
Pupil detection
Refer to
for placement instructions.
9.2.1
Lateral test
Note
•
It is recommended to do the
Lateral
test prior to doing either the
LARP
or
RALP
test. Calibration must be with
Lateral
test type selected.
1. Choose the
Impulse Type
:
Lateral
.
2. For patients with spontaneous nystagmus or gaze-evoked nystagmus (evoked by gaze position during head impulse),
select the check box
Spontaneous Nystagmus
. By selecting this check box for a patient who has spontaneous nys-
tagmus or gaze-evoked nystagmus, it prevents good impulses from being rejected inappropriately.
3. Position the patient at least one meter in front of the wall. (Refer to
for information.)
44
Otometrics - ICS Impulse USB
9 Head Impulse