6. Allowing some flexibility in the cables for head movement during testing, clip the cable clip to the patient’s clothing
at the top of the patient's right shoulder.
7. Ensure the eyes are wide open with eyelids positioned to not interfere with pupil detection. If required, adjust the
skin around the eye:
Tilt the bottom of the goggles out and away from the face, pulling the skin below the eye down and repositioning the
goggles to hold the skin in place. Tilt the top of the goggles out and away from the face, pulling the skin above the eye
up and repositioning the goggles to hold the skin in place.
8. Visually inspect the goggles fit.
Proper goggles fit
7.4
Vision-denied solution and goggles placement
7.4.1
Preparing
1. Do not test in a room with complete darkness. While it may NOT seem intuitive, for optimal results some ambient
room light is beneficial for obtaining a completely light tight test environment for the patient.
2. For patients with long eyelashes, it is advised to remove any mascara. Mascara deposited on the inside of the black
window of the right eye cup (through which the pupil is tracked and the eye is recorded) adversely affects both pupil
tracking and recording of the eye.