Important
•
DO NOT TOUCH THE BLACK WINDOW
When handling the right-eye cup, do not touch the black window. Oils in the skin reduce the effectiveness of the anti-
fog coating on the window. Fingerprints make it difficult to track the pupil.
3. Obtain a vision-denied solution (includes a cup for the right eye and a patch for the left eye).
4. Inform the patient that they will be in complete darkness for a short period. Assure them that, while they cannot see
out, the camera in the goggles can still record their eye movement.
7.4.2
Attaching the right-eye cup
1. Remove the plastic backing from the back side of the cup.
2. Ask the patient to open their eye wide.
Note
•
It is easiest to position the cup if you and the patient are sitting at the same level instead of standing over the
patient.
3. Ask the patient to stare straight ahead. You want the pupil to be in the center of the black window.
4. If the patient’s eyelid droops, hold the eyelid up when placing the cup. This will assist in keeping the patient’s eye
wide open during testing.
5. Fold back the inside edge of the cup placing the ridge against the right side of the nose.
A.
Ridge
B.
Fold back the inside edge
20
Otometrics - ICS Impulse USB
7 Pre-testing setup