DETAILED SETTINGS
8.
Use the LEFT and RIGHT Keys to display
FOCUS on the main display.
Selecting Current (mA) or Voltage (V) Output
9.
Press the UP or DOWN Key.
The sub-display will flash.
10.
Display mA.
Always select the same output as the current/voltage switch selection on the bottom of the Amplifier
Unit.
Setting the First Point (A)
11.
Press the ENT Key.
The display will change to allow the first-point settings to
be made. The output current value will be displayed on
the main display, and the corresponding measured value
will be displayed on the sub-display and the leftmost digit
will flash.
12.
Set the output current value and the corre-
sponding measured value for the first point.
Set a measured value within the measurement distance.
If scaling or calculation has been set, set a value that
reflects those settings.
The flashing digit, i.e., the digit for which a value can be set, will change as shown in the diagram.
SUB
(mm)
POWER
ENABLE
ZERO
Move between digits.
Increment and decre-
ment the numeric value.
SUB
(mm)
POWER
ENABLE
ZERO
SUB
(mm)
POWER
ENABLE
ZERO
SUB
(mm)
POWER
ENABLE
ZERO
Sub-display
digit movement
Digit moves
one at a
time
Movement between main display
and sub-display (from leftmost or
rightmost digit of sub-display)
Summary of Contents for ZX-T -
Page 1: ...OPERATION MANUAL Smart Sensors ZX T Series Cat No E346 E1 04 ...
Page 9: ...6 Preface Contents ZX T Operation Manual PREFACE ...
Page 15: ...12 Preface How to Use This Manual ZX T Operation Manual PREFACE ...
Page 16: ...Section 1 FEATURES 13 ZX T Operation Manual Section 1 FEATURES ZX T Features 14 ...
Page 53: ...50 Section 3 Function Transition Charts ZX T Operation Manual Section 3 BASIC OPERATION ...
Page 103: ...100 Section 5 Error Output Function ZX T Operation Manual Section 5 DETAILED SETTINGS ...
Page 113: ...110 Section 6 Initializing Settings Data ZX T Operation Manual Section 6 AUXILIARY FUNCTIONS ...