Section 4
MAIN APPLICATIONS AND SETTING METHODS
Registering Depth
Use the zero reset function to register the Sensor Head position at the reference depth.
Set a reference sample with a known depth (D) under the Sensor Head.
Switch to FUN mode and set the depth on the Zero Reset display ([ZRDSP]).
Set the offset value and then return to the RUN mode.
Press and hold the ENT Key for at least 1 s to reset to zero.
The Sensor Head position will be registered at the same timing that was used when the
zero reset was executed. (The display value here is 3 mm.) Depth is measured based
on this positional relationship of the Sensor Head.
Refer to
Section 5 Detailed Settings
for details on operation.
Using the Zero Reset Function, p. 92
D
FUN
Offset value setting
Example: D = 3 mm
RUN
Press for 1 s min.
Summary of Contents for ZX-T -
Page 1: ...OPERATION MANUAL Smart Sensors ZX T Series Cat No E346 E1 04 ...
Page 9: ...6 Preface Contents ZX T Operation Manual PREFACE ...
Page 15: ...12 Preface How to Use This Manual ZX T Operation Manual PREFACE ...
Page 16: ...Section 1 FEATURES 13 ZX T Operation Manual Section 1 FEATURES ZX T Features 14 ...
Page 53: ...50 Section 3 Function Transition Charts ZX T Operation Manual Section 3 BASIC OPERATION ...
Page 103: ...100 Section 5 Error Output Function ZX T Operation Manual Section 5 DETAILED SETTINGS ...
Page 113: ...110 Section 6 Initializing Settings Data ZX T Operation Manual Section 6 AUXILIARY FUNCTIONS ...