Zero Reset Operation
4-5
ZS-HL
User’s Manual
Zero Reset Operation
When the zero reset function is used, the reference value “0” is registered as the height
and the measured value can be displayed and output as a positive or negative deviation
(tolerance) from the reference value. In RUN mode, the measured value can be reset to 0
at any timing during measurement.
Example 1: Using the height of sensing object registered as the reference value and the
tolerance output as the measured value
Example 2: Using the height of sensing object as the measured value with an offset set to 10
Example 3: Using zero reset to measure steps in sensing object (zero reset at each
measurement)
0
-2
+2
Sensor Head
(reference value)
Sensing
object
(LOW threshold value)
(HIGH threshold value)
10
+8
+12
Sensor Head
Sensing
object
(LOW threshold value)
(HIGH threshold value)
Sensor Head
Zero reset
executed
This step
is mea-
sured.
Zero reset
executed
Zero reset
executed
Direction of movement
Summary of Contents for ZS--LD10GT
Page 1: ...Smart Sensor 2D CMOS Laser Type Cat No Z236 E1 02A ZS HL Series USER S MANUAL ...
Page 32: ...1 10 Section 1 FEATURES ZS HL User s Manual Section 1 Setting Flow MEMO ...
Page 52: ...2 20 Section2 INSTALLATION CONNECTION ZS HL User s Manual Section2 SmartMonitor ZS MEMO ...
Page 104: ...5 30 Section5 SETTINGS FOR FUNCTIONS ZS HL User s Manual Section5 Clearing the Settings MEMO ...
Page 206: ...9 20 Section9 APPENDIX ZS HL User s Manual Section9 Updating Firmware MEMO ...
Page 210: ...9 24 ZS L User s Manual Section9 Index Section9 APPENDIX MEMO ...