![background image](http://html1.mh-extra.com/html/omron/k6pm-th/k6pm-th_user-manual_743374032.webp)
1 Outline
1 - 14
K6PM-TH Thermal Condition Monitoring Device User’s Manual (H231)
After turning on the power or a software reset, you can check the status of communications with infra-
red thermal sensors on the front panel of the Main Unit.
Changing K6PM-TH Sensor Numbers in Succession
Press the
CH
Key and successively change the K6PM-TH sensor numbers.
*1. After turning on the power or a software reset, up to 30 seconds is required to calculate all temperature data.
*2. "
on
" is displayed for a sensor that sends a response, and "
off
" is displayed for a sensor that sends no
response.
*3. Example: If you select sensor 2 and display the Setting sensor screen in the software tool, "
adj
" will appear.
After the sensor configuration has been registered by using the software tool or communications, the
operation flow is as described below. After turning on the power or a software reset, the Main Unit
enters monitoring mode. The measurement/monitor status can be checked on the front of the Main
Unit.
Changing K6PM-TH Sensor Numbers In Succession
To change K6PM-TH sensor numbers in succession, press the
CH
Key.
Check Segment Values of a Sensor
To change segment numbers in succession, press the
SEG / ALM RST
Key.
The segment numbers change in order from "0" to "15". "99" appears next, indicating sensor internal
temperature.
When the Sensor Configuration Has Not Been Registered ("K6PM-TH
Sensor Search Mode", or "K6PM-TH Sensor Position Adjustment
Mode")
When the Sensor Configuration Has Already Been Registered
("Monitoring Mode")
*1
Number of currently-connected sensors
on
or
off
*2
K6PM-TH sensor search mode
Position adjustment mode
*3
Press
CH
Key
K6PM-TH sensor number
Switch to the
Setting sensor
screen with the
software tool
When the power supply
is turned ON or the
software reset is done.
All lit
Press
CH
Key
Press
CH
Key
Summary of Contents for K6PM-TH
Page 56: ...2 Installation and Wiring 2 20 K6PM TH Thermal Condition Monitoring Device User s Manual H231...
Page 76: ...3 Initial Setting 3 20 K6PM TH Thermal Condition Monitoring Device User s Manual H231...
Page 156: ...7 Troubleshooting 7 8 K6PM TH Thermal Condition Monitoring Device User s Manual H231...
Page 195: ...I 1 I K6PM TH Thermal Condition Monitoring Device User s Manual H231 Index...
Page 197: ......