423
3Z4L Laser Micrometer Protocol
Appendix K
High Calibration Set (Sequence No. 446 (Hex 01BE))
This sequence sets the Laser Micrometer’s high calibration.
Send Data Word Allocation (3rd Operand of PMCR(260))
Allocations are the same as for sequence No. 404 (Calibration Set) except that the LC gauge dimension in
words +4 to +5 are not used.
Receive Data Word Allocation (4th Operand of PMCR(260))
None.
Note 1.
Use sequences No. 446 (High calibration set) and No. 447 (Low calibration set) to calibrate the Laser
Micrometer. If sequence No. 404 (Calibration Set) is used, both the high and low calibration gauges
must be set, and the high calibration and low calibration gauges cannot be exchanged.
2.
The HC gauge dimension must be set with 3 digits for the integer portion and 4 digits for the decimal
portion.
Low Calibration Set (Sequence No. 447 (Hex 01BF))
This sequence sets the Laser Micrometer’s low calibration.
Send Data Word Allocation (3rd Operand of PMCR(260))
Allocations are the same as for sequence No. 404 (Calibration Set) except that the HC gauge dimension in
words +1 to +3 are not used.
Receive Data Word Allocation (4th Operand of PMCR(260))
None.
Note 1.
Use sequences No. 446 (High calibration set) and No. 447 (Low calibration set) to calibrate the Laser
Micrometer. If sequence No. 404 (Calibration Set) is used, both the high and low calibration gauges
must be set, and the high calibration and low calibration gauges cannot be exchanged.
2.
The LC gauge dimension must be set with 3 digits for the integer portion and 4 digits for the decimal
portion.
Summary of Contents for CJ1W-SCU41
Page 3: ...KX ...
Page 5: ...XK ...
Page 9: ......
Page 19: ......
Page 119: ......
Page 177: ......
Page 187: ......
Page 221: ......
Page 259: ......
Page 317: ......
Page 333: ......
Page 373: ......
Page 395: ......
Page 407: ......
Page 443: ......
Page 463: ......
Page 501: ......
Page 509: ......
Page 513: ......
Page 525: ......