173
Error and Warning Messages
201.xxx
Warning [201.xxx] Check connection. Measured ratio > 50000:1.
Reason:
Measurement input "Prim" is possibly not connected properly.
Solution:
Check whether the primary side measurement input "Prim" is connected
properly.
202.xxx
Warning [202.xxx] Check connection. Polarity of input SEC and power output are not
the same.
Reason:
Polarity of input "Sec" and "Output" are not the same.
Solution:
Check whether the output generator ("Output") and measurement input "Sec"
are connected properly and with correct polarity.
203.xxx
Warning [203.xxx] Check connection. Input SEC may be disconnected.
Reason:
Input signal at input "Sec" differs from the expected one.
Solution:
Check whether the output generator ("Output") and measurement input "Sec"
are connected properly.
204.xxx
Warning [204.xxx] Check connection. Polarity of input PRIM and power output are not
the same.
Reason:
Input signal at input "Prim" differs from the expected one: Reverse polarity at
input "Prim".
Solution:
Check whether the output generator ("Output") and measurement input "Prim"
are connected properly and with correct polarity.
204.xxx
Warning [204.xxx] Check CT SB2 switch box wiring: Wrong polarity at PRIM terminal.
Reason:
Input signal at input "PRIM" differs from the expected one: Reverse polarity at
CT SB2
input "PRIM".
Solution:
Check whether the PRIM IN sockets on the
CT SB2
are connected properly
and with correct polarity. If necessary, connect with changed polarity.
205.xxx
Warning [205.xxx] Check connection. Input PRIM may be disconnected.
Reason:
Input signal at input "Prim" differs from the expected one.
Solution:
Check whether the output generator ("Output") and measurement input "Prim"
are connected properly.
205.xxx
Warning [205.xxx] Check CT SB2 switch box wiring: PRIM IN terminal may be
disconnected.
Reason:
Input signal at input "Prim" differs from the expected one.
Solution:
Check whether the PRIM IN sockets on the
CT SB2
are connected properly.
206.xxx
Warning [206.xxx] Check connection. Ensure that the output is connected to the
primary side of the CT.
Reason:
Input signal at input "Prim" differs from the expected one.
Solution:
Check that the output generator ("Output") is connected to the CT and that the
primary side of the CT is connected to input "Prim".
206.xxx
Warning [206.xxx] Check CT SB2 switch box wiring: CTA OUTPUT and/or PRIM OUT
terminal may be disconnected.
Reason:
Input signal at input "Prim" differs from the expected one.
Solution:
Check whether the PRIM OUT and the PRIM IN sockets on the
CT SB2
are
connected correctly to the primary side of the CT.
207.xxx
Warning [207.xxx] Check connection! OUTPUT may be disconnected.
Reason:
The current signal differs from the expected one.
Solution:
Check whether the output generator ("Output") and measurement input "SEC"
are connected properly to the secondary side of the CT.
Summary of Contents for CT Analyzer
Page 1: ...CT Analyzer CT SB2 User Manual...
Page 3: ...CT Analyzer User Manual...
Page 4: ...CT Analyzer User Manual 2 Manual Version CTAnalyzerUM AE 3...
Page 39: ...37 Setup and Connection Figure 2 15 Measurement setup for delta compensation Ratio 1 L2 L3 L1...
Page 50: ...CT Analyzer User Manual 48...
Page 60: ...CT Analyzer User Manual 58...
Page 80: ...CT Analyzer User Manual 78...
Page 162: ...CT Analyzer User Manual 160...
Page 172: ...CT Analyzer User Manual 170...
Page 188: ...CT Analyzer User Manual 186...
Page 194: ...CT Analyzer User Manual 192...
Page 195: ...CT SB2 User Manual...
Page 196: ...CT SB2 User Manual 2 Manual Version CTSB2 AE 3...
Page 204: ...CT SB2 User Manual 10...
Page 206: ...CT SB2 User Manual 12...
Page 211: ...17 Hardware 1 x Cable bag Bag for cables and accessories Accessories Description...
Page 212: ...CT SB2 User Manual 18...
Page 226: ...CT SB2 User Manual 32...
Page 242: ...CT SB2 User Manual 48...
Page 248: ...CT SB2 User Manual 54...
Page 250: ...CT SB2 User Manual 56...
Page 252: ...CT SB2 User Manual 58...